Am J Physiol Lung Cell Mol Physiol 291: L535-L545, 2006. First published May 5, 2006; doi:10.1152/ajplung.00088.2005
1040-0605/06 $8.00
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
291/3/L535    most recent
00088.2005v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Litzlbauer, H. D.
Right arrow Articles by Rau, W. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Litzlbauer, H. D.
Right arrow Articles by Rau, W. S.

INNOVATIVE METHODOLOGY

Three-dimensional imaging and morphometric analysis of alveolar tissue from microfocal X-ray-computed tomography

Horst Detlef Litzlbauer,1,* Christoph Neuhaeuser,2,* Alexander Moell,1 Susanne Greschus,1 Andreas Breithecker,1 Folker Ernst Franke,3 Wolfgang Kummer,4 and Wigbert Stephan Rau1

Departments of 1Diagnostic Radiology, 2Anesthesiology and Critical Care Medicine, 3Institute of Pathology, and 4Institute of Anatomy and Cellular Biology, University of Giessen, Giessen, Germany

Submitted 26 February 2005 ; accepted in final form 6 April 2006

ABSTRACT

We evaluated microfocal X-ray-computed tomography (micro-CT) as a method to visualize lung architecture two and three dimensionally and to obtain morphometric data. Inflated porcine lungs were fixed by formaldehyde ventilation. Tissue samples (8-mm diameter, 10-mm height) were stained with osmium tetroxide, and 400 projection images (1,024 x 1,024 pixel) were obtained. Continuous isometric micro-CT scans (voxel size 9 µm) were acquired to reconstruct two- and three-dimensional images. Tissue samples were sectioned (8-µm thickness) for histological analysis. Alveolar surface density and mean linear intercept were assessed by stereology-based morphometry in micro-CT scans and corresponding histological sections. Furthermore, stereology-based morphometry was compared with morphometric semi-automated micro-CT analysis within the same micro-CT scan. Agreement of methods was assessed by regression and Bland-Altman analysis. Comparing histology with micro-CT, alveolar surface densities (35.4 ± 2.4 vs. 33.4 ± 1.9/mm, P < 0.05) showed a correlation (r = 0.72; P = 0.018) with an agreement of 2 ± 1.6/mm; the mean linear intercept (135.7 ± 14.5 vs. 135.8 ± 15 µm) correlated well (r = 0.97; P < 0.0001) with an agreement of –0.1 ± 3.4 µm. Semi-automated micro-CT analysis resulted in smaller alveolar surface densities (33.4 ± 1.9 vs. 30.5 ± 1/mm; P < 0.01) with a correlation (r = 0.70; P = 0.023) and agreement of 2.9 ± 1.4/mm. Non-destructive micro-CT scanning offers the advantage to visualize the spatial tissue architecture of small lung samples two and three dimensionally.


FOR THREE-DIMENSIONAL structural-functional analysis of pulmonary tissue fixation of the lung and imaging, methods are required that induce minimal structural alterations of the alveolar architecture. The elasticity of the lung and the small amount of deformable tissue organized in a large volume are difficult to master when the native configuration of alveolar parenchyma has to be preserved during preparation (27, 32, 34, 40). The traditional approach to visualize the alveolar space and to perform morphometric measurements is based on histological serial sectioning. Although this method offers high spatial resolution and image contrast, there are important limitations with histological analysis. Information about the three-dimensional pulmonary structure must be reconstructed from calculations of two-dimensional sections (38, 39). Furthermore, the sectioning process itself may result in artifacts. Due to the fragility of the alveolar architecture, pulmonary tissue is prone to collapse, deformation, and disruption during preparation. So far, no histomorphometric method based on a non-destructive three-dimensional alveolar analysis of a continuous data set is available.

An alternate approach to image lung tissue in three dimension is the use of microfocal X-ray-computed tomography (micro-CT) (14, 28, 36). This imaging method provides new opportunities to study the microstructure of intact organs and intact tissue specimens (5, 11, 13, 17). In contrast to histology, with micro-CT scanning, it is possible to investigate tissue specimens non-destructively, without further alteration of the tissue. Continuous digital images are generated for detailed two- and three-dimensional investigation (11, 18, 19, 25, 30). We present an imaging technique that uses micro-CT to visualize small pulmonary structures (e.g., bronchiole, alveolar ducts) from small, prepared air-filled lung specimen. This approach permits three-dimensional micro-CT analysis of the alveolar architecture from inflated lungs. Furthermore, continuous non-destructive data acquisition allows postprocessing of the data for segmentation of bronchial trees and alveolar ducts. To establish micro-CT as a new method for assessment of alveolar architecture, this method was evaluated by comparing the results to the gold standard that still remains: stereology-based histomorphometry (7, 37).

MATERIALS AND METHODS

Isolation of Porcine Lungs

Lungs were harvested from five healthy hybrid pigs (6 mo of age; weight 120 kg) at a local slaughterhouse. No animals were killed for the particular purpose of this study. Only left lungs with intact visceral pleurae were taken for investigation. Lungs were isolated by intersection of the left main bronchus close to the carina. The left main bronchus was subsequently intubated with a cuffed tube and inflated with a breathing bag.

Lung Fixation

Lungs were connected to a tube of a modified Engstrom-type respirator. After recruitment with a peak airway pressure of 40 cmH2O for 40 s, lungs were ventilated in a volume-controlled, pressure-limited mode with a positive end-expiratory pressure of 5 cmH2O. The peak pressure was limited to 35 cmH2O. Tidal volume was adjusted to reach full lung inflation, respiration rate was set at 12/min, and the inspiratory-to-expiratory ratio was 1:2. Fixation occurred by ventilation of the lung with formaldehyde vapor (27). The fixation process lasted 6 h. It resulted in an air-filled lung preparation with inflated alveoli (Fig. 1).


Figure 1
View larger version (85K):
[in this window]
[in a new window]
 
Fig. 1. Radiograph of the lung that reveals homogeneous aeration after fixation with formaldehyde vapor ventilation.

 
Determination of Lung Volume and Weight

To prove the reproducibility of the preparation technique, lung volume and weight were determined before preparation and after fixation.

Lung volume was measured by water displacement (33). Prior preparation volumes were taken in full inflation (near total lung capacity). To avoid deformation of the emerged lung (leading to incorrect values), the intubated main stem bronchus was clamped.

Weight was assessed by use of a common pair of scales.

Assembling of Tissue Specimens

Lungs were cut into parallel slices of 10-mm thickness. Tissue specimens were taken from the second slice counting from the costal surface of the lung. With the use of a sharp mechanical stamp, cylindrical samples (diameter 8 mm, height 10 mm) were gained from two subpleural locations.

Staining of Tissue Specimens

To enhance contrast for micro-CT scanning, tissue samples were immersed in 1% osmium tetroxide solution (1, 4) in tubes. Tubes were rotated gently for 3 h. Afterward, samples were washed threefold for 30 min with purified water to displace excess osmium tetroxide solution in the alveolar air space.

Drying of Stained Tissue Specimens

To remove remaining fluid inside the alveolar air space, the samples were subjected to a critical point drying procedure (15). The samples were incubated threefold with 10 ml of 2,2-dimethoxypropane for 30 min (10).

In the critical point dryer (Critical Point Dryer; Balzers, Witten, Germany), the dimethoxypropane was completely replaced by liquid carbon dioxide. Finally, the chamber was heated to 40°C with a pressure of 90,000 kPa. The critical point of carbon dioxide is reached at a temperature of 31°C and a pressure of 73,800 kPa when carbon dioxide changes abruptly from the liquid to the gaseous phase.

Determination of Diameter and Weight of Tissue Specimen

To test the effect of the preparation technique on the spatial dimensions, the diameter and weight of the tissue specimen were determined before and after osmium staining-critical point drying.

Micro-CT Scanning of Tissue Specimens

Micro-CT scanning of the lung samples was performed on a SkyScan-1072 system (Aartselaar, Belgium). The scanner is equipped with a tungsten microfocus X-ray tube (80 kV type L6731-01 Hamamatsu, Toyooka-village, Japan). It has a beam angle of 39°, reaching a focal spot size of 8 µm. The X-ray detector is a 1,024 x 1,024 12-bit digital charge-coupled device (CCD) camera with a fiber-optic 3.7:1 coupled to a P43 powder scintillator (Gd0) with 3-µm grains. The opacity of each voxel is represented by a 16-bit gray-scale value. Each 16-bit planar image consisted of a 1,024 x 1,024 array of pixels ranging from 0 to 255 (minimum to maximum) X-ray intensity scale.

Scanning was carried out with a tube voltage of 60 kV and a tube current of 100 µA. After performing a scout view, 400 projections of the specimen were scanned over a total rotation of 180° (rotation step was 0.45°). One projection image required 0.3 min, resulting in a total acquisition time of 2 h/sample. Out of the 400 projection images, ~900 continuous axial scans with a matrix of 1,024 x 1,024 were calculated with a modified Feldkamp reconstruction modus (12). According to the cone-beam geometry of the X-ray, the magnification as well as the pixel size were determined by the distance of the object to the X-ray source. The sample diameter of 8 mm permitted a magnification of 27.5, resulting in cubic (isotropic) voxels with a size of 9 µm.

Preparation of Histological Sections For Light Microscopy

After micro-CT scanning, the tissue specimens were embedded in paraffin. From this paraffin block, serial sectioning was performed (135 sections/sample with a thickness of 8 µm). Digital pictures (10-fold lens magnification) were obtained from all intact histological sections using a universal light microscope (Axioplan 2; Zeiss, Jena, Germany) attached to a CCD camera (DXC-3000P, Sony).

Comparison Between Sections From Micro-CT and Histology

The histological slices were matched with the corresponding micro-CT scans for comparison. Slice-to-scan registration of the different imaging modalities was performed with the custom Analyze Software version 4 (Biomedical Imaging Resource; Mayo Foundation, Rochester, MN).

Despite the most possible accuracy, complete agreement between corresponding images could not be accomplished within micrometers due to angular mismatching in the specimen orientation. Histological sections did not correspond exactly to the transversal micro-CT slices. Thus matching layers were extracted on the basis of anatomical structures (e.g., greater bronchiole, interlobular septae) with a three-dimensional tool providing the facility to generate arbitrary orientated slices of the entire volume (Fig. 2, A and B).


Figure 2
View larger version (71K):
[in this window]
[in a new window]
 
Fig. 2. Microfocal X-ray-computed tomography (micro-CT) scan (A) and corresponding histological section (B). Groove for adjustment at 6 o’clock (*). Note the straight orientation of the septae in micro-CT in contrast to the wavy appearance in histology. The section of the airway at 12 o’clock appears round in micro-CT and slightly deformed in histology.

 
Three-Dimensional Postprocessing of Two-Dimensional Micro-CT Scans

Three-dimensional postprocessing of the images was performed with the custom ANALYZE software (Mayo Foundation). Two-dimensional, eight-bit bitmap files were transformed into three-dimensional volumes and visualized by volume rendering (Fig. 3, A and B).


Figure 3
View larger version (107K):
[in this window]
[in a new window]
 
Fig. 3. The continuous, isotropic data set of micro-CT allows the reconstruction of 3-dimensional images. Because of the acquisition of isovolumetric voxels, there is no loss of resolution in different planes, axial (A) and coronal (B) orientation.

 
Segmentation of Bronchiolar Tree and Pulmonary Acinus

The three-dimensional structure of the conductive airway tree was reconstructed with a seeding technique, considering equal Hounsfield units and connectivity. The seed was placed in a bronchus, and automated segmentation was carried out under visual control of the growing seed. Thus segmentation was performed in a process that was based on an automated algorithm in addition to examiner supervision and guidance. Following the bronchiolar tree, the seeding technique was performed again starting at the terminal bronchus. The airways were followed down to the alveolar walls (Fig. 4, A and B). Visualization of the segmented air space unit in a surface three-dimensional rendering is demonstrated in Fig. 4, C and D. The software employed in this step was Volume Tool of the Advantage Workstation 4.1 (General Electric Medical Systems Europe, Buc, France).


Figure 4
View larger version (124K):
[in this window]
[in a new window]
 
Fig. 4. Segmentation of the pulmonary acinus based on a seeding technique. Coronal (A) and transverse (B) scan of the original data of the segmentation. First, the conductive airway (arrows) was segmented, placing a seed in a bronchus. The starting point of the second segmentation was the last bronchiole without alveoli (arrowhead). Distal to this point, the airways were segmented down to the alveolar walls (the red line indicates the border of the acinus). Surface rendering of the resulting unit is shown from two perpendicular projections (C and D).

 
Morphometric Data Determined By Semiautomated Micro-CT Analysis

We chose alveolar surface density and the mean linear intercept (an estimate for the alveolar diameter) as morphometric parameters for comparison with stereology-based morphometry.

For semiautomated micro-CT analysis, a cubic volume of interest (side length 702 µm including 78 continuous scans; volume of interest = 0.346 mm3) had to be defined within the reconstructed micro-CT images of the samples that just covered alveoli (large bronchi and vessels were excluded by visual control; see Fig. 6A). The gray-scale images were transformed into binary images using a threshold plug-in according to the Canny method (8) (see Fig. 6, C and D). Morphometric analysis was performed with the custom ANALYZE software. The alveolar (tissue) volume and the alveolar surface area can be assessed directly from micro-CT scans; the alveolar surface density can be calculated from these morphometric parameters.

Formula 1(1)
The mean linear intercept was calculated by the general equation

Formula 2(2)
where k is a factor with the value 6 (see APPENDIX) and alveolar air space volume is the difference between the volume of interest and alveolar tissue volume.


Figure 6
View larger version (104K):
[in this window]
[in a new window]
 
Fig. 6. For morphometric micro-CT analysis, a region of interest (ROI) was defined that covered only alveoli (square in A). B: ROI (side length 700 µm) of the gray-scale value image with the grid of Merz superimposed. Arrows indicate the linear intercepts. C: red line shows the border of the threshold-based segmentation of tissue and air. D: binary image. With threshold-based segmentation, some of the tissue structures are not included (B and D).

 
Histological Data Determined By Stereology-Based Morphometry

For histomorphometry, the digital images acquired during the histological examination were processed by an IBAS 2 image processor (Kontron, Germany). A similar region of interest as previously defined in the micro-CT measurement was analyzed.

Alveolar Surface Density

According to a fundamental equation of stereology originally obtained by Saltykov (31), the surface density (surface area ÷ tissue volume) can be estimated by 2x (number of intersections ÷ length of test lines) (7, 31).

An overlay image providing a measuring grid according to Merz (24) with a side length of 700 µm was used to count the number of intercepts (Fig. 5).


Figure 5
View larger version (181K):
[in this window]
[in a new window]
 
Fig. 5. Histological slice, the grid of Merz (side length 700 µm) superimposed. The number of intercepts was used for estimation of the surface density. Arrows indicate the measurement of the linear intercepts.

 
The total length of test lines of the Merz grid is 3.5 x {pi} mm (see APPENDIX).

Thus the surface density can be estimated by the equation

Formula 3(3)
The number of intercepts was evaluated in 20 grids of Merz per sample (2 grids/section x 10 sections).

Mean Linear Intercept

An overlay image providing a measuring grid according to Merz (24) was used to define intercepts with the alveolar borders (Fig. 5). These intercepts were used to define the starting point for a two-point measurement of the intercept length. In total, 125 intercept lengths per sample were assessed (5 lengths/set x 5 sets/section x 5 sections). Only completely preserved closed alveolar structures were considered.

Stereology-Based Morphometric Data From Micro-CT

For micro-CT morphometry, the same grid of Merz was overlayed on the micro-CT scans.

Alveolar Surface Density

The surface density was calculated according to Eq. 3.

The number of intercepts was evaluated in 20 grids of Merz per sample (2 grids/scan x 10 scans; Fig. 6B).

Mean Linear Intercept

The grid of Merz was used for a two-point measurement of the intercept length. In total, 125 lengths per sample were assessed by the two-point measurement described above (Fig. 6B).

Statistics

Data were tested by analysis of variance. Correlation was tested by bivariate regression analysis. P < 0.05 was considered as significant.

The agreement between alveolar surface density and the mean linear intercept obtained by semiautomated micro-CT analysis and stereology-based morphometry within the same micro-CT scans was assessed according to Bland and Altman (6). In addition, the agreement of the stereology-based morphometric data assessed in micro-CT scans and in histological slices was tested.

RESULTS

Volumes and Weights of Whole Lungs Before and After Preparation

The mean volume of inflated lungs decreased only by 3% during formaldehyde vapor ventilation. Mean lung volumes did not differ significantly before and after lung fixation (2,107 ± 309 ml vs. 2,047 ± 296 ml). In contrast, the mean lung weight decreased by 70% during fixation (364 ± 27 g vs. 109 ± 9 g; P < 0.05).

Dimensions of Tissue Specimen Before and After Staining and Drying

Further preparation of the cylindrical lung tissue specimens by osmium tetroxide staining and critical point drying resulted in a sample shrinkage of 10.4% regarding the sample diameter and further weight loss of 37.1% in average (Table 1).


View this table:
[in this window]
[in a new window]
 
Table 1. Diameter and weights of cylindrical tissue samples before and after osmium tetroxide staining and critical point drying

 
Comparison of Micro-CT Scans and Histological Sections

Micro-CT scanning resulted in an isotropic voxel size of 9 µm with a resolution of 18 µm. Realistic two- and three-dimensional images of the sample could be reconstructed.

The comparison of micro-CT scans to corresponding histological sections of the same sample showed a high agreement by visual control (Fig. 2). However, exact matching is not always possible on a microscopic level.

Basic Descriptive Statistic of Histological and Micro-Tomographic Parameters

Semiautomated micro-CT analysis. Air space volume and alveolar surface (primary morphometric parameters) assessed by micro-CT analysis are shown in Table 2.


View this table:
[in this window]
[in a new window]
 
Table 2. Primary morphometric parameters assessed by micro-CT analysis

 
Column three in Table 2 is used to calculate the mean linear intercept (APPENDIX).

Comparison of Stereology-Based Results Between Micro-CT and Histology

To evaluate the accuracy of micro-CT, stereology-based morphometric measurements were compared between histological images and micro-CT scans.

Alveolar surface density. Stereology-based measurements of the alveolar surface density (Table 3) differed significantly between histological and micro-CT images (35.47 ± 2.41/mm vs. 33.47 ± 1.91/mm). The regression analysis (Fig. 7A) showed a correlation of r = 0.72 (P = 0.018); the agreement was 2 ± 1.6/mm in the Bland-Altman plot (Fig. 7B).


View this table:
[in this window]
[in a new window]
 
Table 3. Alveolar surface density assessed by micro-CT morphometric analysis and by stereology-based measurement on micro-CT scans as well as on histological slices

 

Figure 7
View larger version (22K):
[in this window]
[in a new window]
 
Fig. 7. Alveolar surface density: regression analysis and Bland-Altman plots for the comparison between methods are shown. A and B: comparison of stereology-based measurements between histological and micro-CT images. C and D: comparison between stereology-based measurements and semiautomated micro-CT analysis within the same micro-CT scan.

 
Mean linear intercept. In contrast to alveolar surface density, stereology-based measurements of the mean linear intercept (Table 4) did not differ between histological and micro-CT images (135.7 ± 14.5 µm vs. 135.8 ± 15 µm). The correlation of mean linear intercepts was r = 0.97 (P < 0.0001), with an agreement of –0.1 ± 3.4 µm (Fig. 8, A and B).


View this table:
[in this window]
[in a new window]
 
Table 4. Mean linear intercept assessed by micro-CT morphometric analysis and by stereology-based measurement on micro-CT scans as well as on histological slices

 

Figure 8
View larger version (22K):
[in this window]
[in a new window]
 
Fig. 8. Mean linear intercept: regression analysis and Bland-Altman plots for the comparison between methods are shown. A and B: comparison of stereology-based measurements between histological and micro-CT images. C and D: comparison between stereology-based measurements and semiautomated micro-CT analysis within the same micro-CT scan.

 
Comparison Between Semiautomated and Stereology-Based Results Within the Same Micro-CT Image

To evaluate the accuracy of semiautomated micro-CT morphometric analysis, stereology-based measurements were compared with semiautomated micro-CT analysis within the same micro-CT scan.

Alveolar surface density. The alveolar surface density (Table 3) differed significantly between stereology-based measurement and semiautomated analysis within the same micro-CT volume of interest (33.47 ± 1.91/mm vs. 30.49 ± 0.95/mm). The regression analysis (Fig. 7C) showed a correlation of r = 0.7 (P = 0.023); the agreement was 2.9 ± 1.4/mm in the Bland-Altman plot (Fig. 7D).

Mean linear intercept. The mean linear intercept (Table 4), calculated by Eq. 2, did not differ significantly from mean linear intercept assessed by stereology-based morphometry (134.4 ± 13.06 µm vs. 135.8 ± 15 µm). The regression analysis showed a correlation of r = 0.97 (P < 0.0001); the agreement was 1.4 ± 3.3 µm (Fig. 8, C and D).

DISCUSSION

The classic approach to image alveolar tissue is based on histological serial sectioning. A major limitation of histology is the destructive nature of preparation. Even adjacent slices often do not match exactly due to tissue deformation. In contrast, micro-CT offers the possibility to image tissue non-destructively by creating a continuous data set of the investigated object (13). However, alveolar architecture, in contrast to cancellous bone (20, 30), is rather difficult to study with micro-CT because of the low spatial stability and the inadequate intrinsic contrast of native lung tissue.

To solve this problem, preparation techniques must be applied that avoid alveolar collapse and furthermore enhance radiodensity. Various methods of preparing inflated lungs have been described; some employ vascular perfusion with fixatives (2), others employ rapid freezing followed by permeation with fixatives (23) or formaldehyde inflation (27, 32, 40). In our study, formaldehyde ventilation resulted in stable air-filled lungs with a mean volume of 2 l for the left lung. In adult pigs, total lung capacity was reported to be 4.3 l in average (35). However, there is no question that formaldehyde fixation alters lung tissue, despite constant total lung volumes in our study. Proteins (e.g., collagen, elastin) are denaturated and the tissue is dehydrated. As a consequence, lung tissue loses elasticity, and an average volume loss of up to 20% was described for formaldehyde fixation (22). Differences between native and fixed lung volumes must be considered for morphometric analysis (41). In addition, lung preparations should be timely analyzed to avoid further changes by shrinkage over time (5). Another important issue is the constancy of alveolar volumes during preparation. Whether alveolar volumes behave comparable to total lung volume during fixation cannot be answered with certainty. Most investigators believe that alveoli unfold in the tidal breathing range and only become stretched at high volume (3). However, others believe that the alveolar volume has a rather persistent dimension. Using in vivo microscopy, Carney et al. (9) showed that the increase in lung volume from 20 to 80% of total lung capacity was based mainly on recruitment of previously deflated alveoli and was not accompanied by a significant change in the individual alveolar volume.

The main problem regarding micro-CT analysis of pulmonary microstructure is the inadequate intrinsic radiodensity of native lung tissue that prevents imaging with high resolution. Whereas very small samples of native tissue (~1 mm3) may be imaged, larger samples of unstained preparations do not have sufficient contrast between the thin alveolar walls and the air space for satisfying data acquisition. Initial attempts were made using silver nitrate as dye followed by gentle centrifugation to remove the staining solution from the air space (36). This procedure did not succeed in a homogeneous staining of the tissue. The centrifugation resulted in deformation of the alveolar architecture that interfered with morphometric analysis. Therefore, we used osmium tetroxide as dye and fixative (4). Osmium tetroxide binds covalently to unsaturated fatty acids of the cell membrane. This results in further tissue stabilization. Due to the high atomic number of 76, it offers excellent radiodensity. The most critical step in our preparation was to eliminate excessive osmium tetroxide solution from the air spaces. The best results were achieved using a critical point drying procedure. With this technique, effects of surface tension are avoided, and the entire fluid is eliminated without tissue deformation. However, some extent of shrinkage does occur (22). A total volume loss of about 10–14% in relation to the native lungs must therefore be taken into consideration. To calculate native lung parameters (e.g., mean linear intercept), a correction factor can be calculated from the relation of sample diameters before and after critical point drying (8.17 ÷ 7.32 = 1.11).

We achieved two- and three-dimensional micro-CT images of alveolar tissue with a voxel size of 9 µm and resolution of 18 µm. Corresponding micro-CT scans and histological images showed an excellent matching of structures by visual control (Fig. 2, A and B). Perfect congruence on a microscopic level, however, is not feasible because of angular distortions between methods. In addition, it must be addressed that classification of tissues (e.g., cell types), which can be easily done by histology, is impossible with micro-CT.

One of the goals of this study was to establish micro-CT as a method for morphometric investigation of alveolar tissue. Stereology-based histological analysis of perpendicular slices can be considered as the gold standard for morphometric measurements (7, 38). To evaluate the accuracy of micro-CT imaging (transforming the different X-ray attenuations of the tissue into gray-value images), alveolar surface density measured by a modified stereology-based method within corresponding histological sections and micro-CT images was compared. Micro-CT resulted in consistently lower values with good correlation, however. This may be explained by the limited resolution of micro-CT analysis. Because alveolar septae are only 5–10 µm thick, which lies in the range of one voxel (9 µm), very thin walls are not sufficiently represented in micro-CT images.

In addition, we evaluated the accuracy of a semiautomated micro-CT analysis to obtain morphometric data. Alveolar surface density determined by a modified stereology-based method and semiautomated micro-CT analysis was compared within the same micro-CT scan. Semiautomated micro-CT analysis resulted in reasonable data, but the values were lower compared with the stereology-based measurements in micro-CT scans and histology, respectively. The critical step in quantitative assessment of semiautomated micro-CT analysis is to create a binary model of the tissue by using a threshold procedure. Although the tissue/air contrast of our preparation is very high (10:1), care must be taken that thick walls are not overrepresented while thin walls are lost using a global threshold procedure (Fig. 6, C and D). Feldkamp et al. (13) and Kuhn et al. (20) addressed this issue by pointing out that X-rays within heterogeneous material are inconsistently attenuated, not only between different specimens but also within the same sample. We set the threshold level by comparing the binary images with the histological slices. However, if the threshold level chosen is too high, alveolar surface density will be too large, whereas mean linear intercept will be too small, and vice versa. We believe that the smaller values for the alveolar surface density are due to loss of thin walls during transformation into binary images. To date, the resolution of histology is not achieved with micro-CT. Although the gray-scale value scans look very similar compared with the histological slices, the binary images do not. With a voxel size of 9 µm, a realistic binary tissue model is not feasible. In spite of the difference in attenuation between alveolar tissue and air, the thickness of alveolar walls is not sufficient enough to create a realistic binary tissue model. To some extent, this can be explained by the partial volume effect of CT imaging. Because the voxel size has a finite dimension, an error occurs if more attenuations than one (e.g., air and tissue) are present within a voxel. The most significant limitation of the threshold segmentation is to assume that each voxel contains one type of tissue. This segmentation is incompatible with volume averaging as voxels representing different types of X-ray attenuation cannot be classified correctly. The effect of volume averaging is greatest at tissue-surface interfaces. This geometric reality makes the accurate assessment of surfaces by means of threshold segmentation difficult. The effect of volume averaging can be reduced if the voxel size is decreased. The use of synchrotron radiation sources permits a high spatial resolution down to the micrometer level (26). With these advanced techniques, morphometric analysis may be enhanced.

In contrast to alveolar surface density, agreement for mean linear intercept as an estimate for the alveolar diameter was much better. This finding can be explained by the fact that the diameter is not as sensitive for limitations in resolution. Based on histomorphometric investigation, Lum and Mitzner (21) determined an alveolar diameter of 133 µm in pig. Their data are consistent with our results. Even semiautomated micro-CT analysis, which uses primary data as volume and surface for the calculation of alveolar diameter, resulted in reasonable data. However, a proportionality factor of 6 had to be incorporated into the formula (see APPENDIX).

It should be noted that the morphometric micro-CT analysis is not a fully automated procedure. Micro-CT morphometry still depends on human judgment, since the program does not distinguish between anatomical structures. The inclusion of bronchi or vessels in the volume of interest will result in errors in alveolar morphometric data.

The advantage of micro-CT is that relying on the isotropic data set imaging of three-dimensional structures like bronchioli and acini can be easily performed. The traditional anatomical concept holds that the acinus begins at the last airway without alveoli (the transitional bronchiole) (16). All airways proximal to the transitional bronchiole are purely of conducting nature. Segmentation of the acinus can be done by tracking down a terminal bronchiole to the alveolar air space. The segmentation process started in placing a seed in a section of a bronchus. Under visual control of the growing seed, a model of the conductive airway was constructed. As shown in Fig. 4A, there was a clear transition in the wall thickness of the conductive airways and the alveolated airways. Thus defining a transitional bronchiole (the last airway without alveoli) as the starting point for the segmentation of the acinus was reliable. In a second segmentation, the airways were followed down distally to the alveolar walls, resulting in the visualization of the acinus (Fig. 4, C and D). To validate the correct calculation, the borders of the segmentation (Fig. 4, A and B) were checked in the two-dimensional scans by the investigator. Therefore, the segmentation still depends on human judgment. To optimize the segmentation, further work needs to be done to improve the resolution of the micro-CT scanning process.

In conclusion, we showed that semiautomated micro-CT analysis is feasible in fixed and radiological-enhanced lung tissue samples. However, the accuracy of morphometric measurements is restricted to the applied voxel size that determines resolution. Because thickness of alveolar septae is only 5–10 µm, a voxel size <3 µm is required for precise tissue quantification.

In contrast to histology, the isotropic continuous data set allows the reconstruction of three-dimensional images and may be used for segmentation of the pulmonary acinus. As the preparation technique can be applied in human lungs, there may be an interesting field for morphometric investigations of human pulmonary pathologies (e.g., fibrosis, emphysema, or pneumoconiosis) performed on postmortem lungs.

APPENDIX

Grid of Merz: Length of Test Line

The grid of Merz is composed of ten lines. Each line is composed of 10 semicircles (corresponding 5 circles) with the diameter (d).

The d of one circle is the tenth part of the side length (700 µm) of the square (d = 70 µm). The circumference of a circle is {pi} x 70 µm. The total length of the test lines:

Formula A1(A1)

Alveolar Diameter

Under the assumption that alveoli are ideal spheres, d of a single alveolus can be calculated when the air space volume and the surface of the alveoli are known.

Formula A2(A2)

Formula A3(A3)
The d can be calculated by:

Formula A4(A4)

Formula A5(A5)
By inserting Eqs. A2 and A3 into Eq. A5:

Formula A6(A6)
In this case the proportionality factor (k) gives the number 6.

Under the assumption that the alveoli in our volume of interest can be considered as a group of spheres with different diameters, the situation is more complex. k depends not only on the distribution of the spheres but also on the variation of individual diameters.

Homogeneous Distribution of Spheres: Example 1: Different Diameters

In the case of five small spheres (d = 1 mm) and five large spheres (D = 2 mm), the mean diameter (dmean = 1.5 mm) can be calculated by Eq. A4.

The air space volume (V) is the sum of the volume of the small spheres [Vd = 5 x ({pi} x d3) / 6] and the volume of the five large spheres [VD = 5 x ({pi} x D3) / 6]:

Formula A7(A7)

Formula A8(A8)
The surface area is the sum of the area of the five small spheres (Ad = 5 x {pi} x d2) and the area of the five large spheres (AD = 5 x {pi} x D2):

Formula A9(A9)

Formula A10(A10)
By inserting Eqs. A8 and A10 into Eq. A4:

Formula A11(A11)
In this example, k is 5.

Example 2: Similar Diameters

In the case of five small spheres (d = 1 mm) and five large spheres (D = 1.1 mm), the mean diameter (dmean = 1.05 mm) can be calculated by Eq. A4.

By inserting Eqs. A8 and A10 into Eq. A4:

Formula A12(A12)
In this example, k is 5.97.

Heterogeneous Distribution of Spheres: Example 3: Different Diameters

In the case of one small sphere (d = 1 mm) and nine large spheres (D = 2 mm), the mean diameter (dmean = 1.9 mm) can be calculated by Eq. A4.

Analog to Eq. A7 the sum of the volumes is:

Formula A13(A13)
Analog to Eq. A9 the sum of the areas is:

Formula A14(A14)
By inserting Eqs. A13 and A14 into Eq. A4:

Formula A15(A15)
In this example, k is 5.78.

If the variation of the d is small (e.g., <10%), k reaches 6 regardless of the distribution of the spheres.

ACKNOWLEDGMENTS

We thank Gerd Magdowski for expert technical advice in osmium tetroxide staining and critical point drying of the tissue samples. We are grateful to Gunhild Martels for micro-CT scanning and Tamara Papadakis for histological preparation.

FOOTNOTES


Address for reprint requests and other correspondence: H. D. Litzlbauer, Klinikstrasse 29, D-35392 Giessen, Germany (e-mail: hdlitz{at}yahoo.de)

The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.

* H. D. Litzlbauer and C. Neuhaeuser contributed equally to this work. Back

REFERENCES

  1. Adams CWM. Osmium tetroxide and the Marchi method: reactions with polar and non-polar lipids, proteins and polysaccharide. J Histochem Cytochem 8: 262–267, 1960.[Abstract]
  2. Bachofen H, Ammann A, Wangensteen D, and Weibel ER. Perfusion fixation of lungs for structure-function analysis: credits and limitations. J Appl Physiol 53: 528–553, 1982.[Abstract/Free Full Text]
  3. Bachofen H, Schürch S, Urbinelli M, and Weibel ER. Relations among alveolar surface tension, surface area, volume, and recoil pressure. J Appl Physiol 62: 1878–1887, 1987.[Abstract/Free Full Text]
  4. Behrmann EJ. The chemistry of osmium tetroxide fixation. In: Science of Biological Specimen Preparation, edited by AMF O’Hare. Chicago: Scanning Electron Microscopy, 1983, p. 1–5.
  5. Bentley MD, Ortiz MC, Ritman EL, and Romero JC. The use of microcomputed tomography to study microvasculature in small rodents. Am J Physiol Regul Integr Comp Physiol 282: R1267–R1279, 2002.[Abstract/Free Full Text]
  6. Bland JM and Altman DG. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet 346: 1085–1087, 1995.[CrossRef][ISI][Medline]
  7. Bolender RP, Hyde DM, and Dehoff RT. Lung morphometry: a new generation of tools and experiments of organ, tissue, cell, and molecular biology. Am J Physiol Lung Cell Mol Physiol 265: L521–L548, 1993.[Abstract/Free Full Text]
  8. Canny JF. A computational approach to edge detection. IEEE Trans Pattern Anal 185: 679–698, 1986.
  9. Carney DE, Bredenberg CE, Schiller HJ, Picone AL, McCann UGII, Gatto LA, Bailey G, Fillinger M, and Nieman GF. The mechanism of lung volume change during mechanical ventilation. Am J Respir Crit Care Med 160: 1697–1702, 1999.[ISI][Medline]
  10. Conway K and Kiernan JA. Chemical dehydration of specimens with 2,2-dimethoxypropane (DMP) for paraffin processing of animal tissues: practical and economic advantages over dehydration in ethanol. Biotech Histochem 74: 20–26, 1999.[ISI][Medline]
  11. Duvall CL, Taylor WR, Weiss D, and Guldberg RE. Quantitative microcomputed tomography analysis of collateral vessel development after ischemic injury. Am J Physiol Heart Circ Physiol 287: H302–H310, 2004.[Abstract/Free Full Text]
  12. Feldkamp LA. Practical cone-beam algorithm. J Opt Soc Am 1: 612–619, 1984.
  13. Feldkamp LA, Goldstein SA, Parfitt AM, Jesion G, and Kleerekoper M. The direct examination of three-dimensional bone architecture in vitro by computed tomography. J Bone Miner Res 4: 3–11, 1989.[ISI][Medline]
  14. Flannery BP, Deckman WG, Roberge WG, and D’Amico KL. Three dimensional X-ray microtomography. Science 273: 1439–1444, 1987.
  15. Gusnard D and Kirschner RH. Cell and organelle shrinkage during preparation for scanning electron microscopy: effects of fixation, dehydration and critical point drying. J Microsc 110: 51–57, 1977.[ISI][Medline]
  16. Haefeli-Bleuer B and Weibel ER. Morphometry of the human pulmonary acinus. Anat Rec 220: 401–414, 1988.[CrossRef][Medline]
  17. Jorgensen SM, Demirkaya O, and Ritman EL. Three-dimensional imaging of vasculature and parenchyma in intact rodent organs with X-ray micro-CT. Am J Physiol Heart Circ Physiol 275: H1103–H1114, 1998.[Abstract/Free Full Text]
  18. Karau KL, Johnson RH, Molthen RC, Dhyani AH, Haworth ST, Hanger CC, Roerig DL, and Dawson CA. Microfocal X-ray CT imaging and pulmonary arterial distensibility in excised rat lungs. Am J Physiol Heart Circ Physiol 281: H1447–H1457, 2001.[Abstract/Free Full Text]
  19. Karau KL, Molthen RC, Dhyani A, Haworth ST, Hanger CC, Roerig DL, Johnson RH, and Dawson CA. Pulmonary arterial morphometry from microfocal X-ray computed tomography. Am J Physiol Heart Circ Physiol 281: H2747–H2756, 2001.[Abstract/Free Full Text]
  20. Kuhn JL, Goldstein SA, Feldkamp LA, Goulet RW, and Jesion G. Evaluation of a microcomputed tomography system to study trabecular bone structure. J Orthop Res 8: 833–842, 1990.[CrossRef][ISI][Medline]
  21. Lum H and Mitzner W. A species comparison of alveolar size and surface forces. J Appl Physiol 62: 1865–1871, 1987.[Abstract/Free Full Text]
  22. Lum H and Mitzner W. Effects of 10% formalin fixation on fixed lung volume and lung tissue shrinkage. A comparison of eleven laboratory species. Am Rev Respir Dis 132: 1078–1083, 1985.[ISI][Medline]
  23. Mazzone RW, Durand CM, and West JB. Electron microscopy of lung rapidly frozen under controlled physiological conditions. J Appl Physiol 45: 325–333, 1978.[Abstract/Free Full Text]
  24. Merz WA. Die Streckenmessung an gerichteten Strukturen im Mikroskop und ihre Anwendung zur Bestimmung von Oberflaechen-Volumen-Relationen im Knochengewebe. Mikroskopie 22: 132–142, 1968.[Medline]
  25. Muller R, Van Campenhout H, Van Damme B, Van Der Perre G, Dequeker J, Hildebrand T, and Ruegsegger P. Morphometric analysis of human bone biopsie: a quantitative structural comparison of histological sections and micro-computed tomography. Bone 23: 59–66, 1998.[Medline]
  26. Peyrin F, Salome M, Nuzzo S, Cloetens P, Laval-Jeantet AM, and Baruchel J. Perspectives in three-dimensional analysis of bone samples using synchrotron radiation microtomography. Cell Mol Biol 46: 1089–1102, 2000.[ISI][Medline]
  27. Rau WS, Hauenstein K, Mittermayer C, and Volk P. A simple and rapid method for postmortem radiographic investigation of lung fine structure. Path Res Pract 170: 426–432, 1980.[Medline]
  28. Ritman EL. Micro-computed tomography: current status and developments. Annu Rev Biomed Eng 6: 185–208, 2004.[CrossRef][ISI][Medline]
  29. Rodriguez-Porcel M, Lerman A, Ritman EL, Wilson SH, Best PJM, and Lerman LO. Altered myocardial microvascular 3D architecture in experimental hypercholesterolemia. Circulation 102: 2028–2030, 2000.[Abstract/Free Full Text]
  30. Ruegsegger P, Koller B, and Müller R. A microtomographic system for the nondestructive evaluation of bone architecture. Calcif Tissue Int 58: 24–29, 1996.[ISI][Medline]
  31. Saltykov SA. The method of intersections in metallography. Zavodskaja Lab 12: 816–825, 1946.
  32. Satoh K, Kobayashi T, Ohkawa M, and Tanabe M. Preparation of human whole lungs inflated and fixed for radiologic-pathologic correlation. Acad Radiol 4: 374–379, 1997.[CrossRef][ISI][Medline]
  33. Scherle WA. A simple method for volumetry of organs in quantitative stereology. Mikroskopie 26: 57–60, 1970.[Medline]
  34. Stockinger L and Hofer H. Preparation and fixation dependence of alveolar structures. Anat Anz 127: 200–209, 1970.[Medline]
  35. Tenney SM and Remmers JE. Comparative quantitative morphology of the mammalian lung: diffusing area. Nature 197: 54–56, 1963.[CrossRef][Medline]
  36. Watz H, Breithecker A, Rau WS, and Kriete A. Micro-CT of the human lung: imaging of alveoli and virtual endoscopy of an alveolar duct in a normal lung and in a lung with centrilobular emphysema - initial observations. Radiology 236: 1053–1058, 2005.[Abstract/Free Full Text]
  37. Weibel ER. Measuring through the microscope: development and evolution of stereological methods. J Microsc 155: 393–403, 1989.[ISI][Medline]
  38. Weibel ER. Morphometry of the human lung: the state of the art after two decades. Bull Eur Physiopath Respir 15: 999–1013, 1979.[ISI][Medline]
  39. Weibel ER and Gomez DM. Architecture of the human lung. Science 137: 577–587, 1962.[Abstract/Free Full Text]
  40. Weibel ER and Vidone RA. Fixation of the lung by formalin steam in a controlled state of air inflation. Am Rev Respir Dis 84: 856–861, 1961.[ISI][Medline]
  41. Xiao Y, Carbayo JJP, Weibel ER, and Hsia CCW. Variation of lung volume after fixation when measured by immersion or Cavalieri method. Am J Physiol Lung Cell Mol Physiol 284: L242–L245, 2003.[Abstract/Free Full Text]




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
291/3/L535    most recent
00088.2005v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Litzlbauer, H. D.
Right arrow Articles by Rau, W. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Litzlbauer, H. D.
Right arrow Articles by Rau, W. S.